We take your personal privacy very seriously. When you visit our website, please agree to all the cookies used. For more information on the processing of personal data, please refer to our Privacy terms

ADX923AQFN32

  • Integrated respiration impedance, three low-noise PGAs, and two 24-bit high-precision ADCs for analog ECG front-end.

Product details

  • Parameters

  • Characteristics

  • Package

  • Description

    • Integrated respiration impedance, three PGAs, two 24-bit high-precision ADCs, SPI communication, and a sampling rate of 16Ksps.

    • -40°C to+85°C

    • •Three low-noise PGA (ADX923) and two high-resolution ADCs (ADX922/3).

    • • Input-referred noise.

      8.6μVPP (G = 6) 

    • • Input bias current

      100pA 

    • • Data rate

      125SPS 至 8kSPS 

    • • Common Mode Rejection Ratio

      117dB

    • • Programmable gain

      1, 2, 3, 4, 6, 8, or 12

    • Analog power supply

      2.7V to 5.5V

    • digital power supply

      1.65V to 3.6V

    • • Built-in right leg drive amplifier, lead-off detection, test signal.

    • • Integrated respiratory impedance measurement (ADX922/3).

    • • Electrode Rotation Mode (ERM) enables time-division multiplexing for LEAD-I, LEAD-II, and LEAD-III measurements (ADX923).

    • • Integrated simulation PACE measurement (ADX921/2/3)

    • High speed mode (ADX921/2/3)

    • • Built-in Oscillator and Reference

    • • Power supply monitor, internal temperature sensor, OSC failure detection, over-range detection.

    • • Flexible power-down and standby modes

    • • SPI™ compatible serial interface with timeout reset functionality

    • • Communication with cyclic redundancy check (CRC) and checksum

    • • Supports ADC data FIFO to reduce polling time for external devices and conserve power

    • • Functional safety

    • • Supports systems compliant with IEC 60601-2-47 standard  • Operating temperature range

      –40°C to 85°C 

    • QFN32 4mm x 4mm

  • The ADX920/1/2/3 devices are multi-channel, simultaneous sampling, 24-bit, Delta-Sigma (ΔΣ) analog-to-digital converters (ADCs) that feature a programmable gain amplifier (PGA), an internal reference, and an on-board oscillator. These devices integrate all the functionalities typically required for portable, low-power medical electrocardiogram (ECG) applications, as well as for fitness and exercise monitoring systems. With their high level of integration and excellent performance, the ADX920/1/2/3 devices enable the creation of scalable medical instrument systems with significantly reduced size, power consumption, and overall cost. Each channel of the ADX920/1/2/3 devices includes a flexible input multiplexer that can independently connect to internally generated signals for testing, temperature measurement, and lead-off detection. Moreover, any input channel configuration can be configured to derive a right-leg drive (RLD) output signal. The ADX920/1/2/3 devices operate at data rates up to 8 kSPS. Lead-off detection can be accomplished internally using the device's current sink or source excitation capabilities. The ADX922/3 models additionally feature fully integrated respiratory impedance measurement functionality. The specified operating temperature range is from –40°C to 85°C.

Related applications

Do you need to inquire more product information?