We take your personal privacy very seriously. When you visit our website, please agree to all the cookies used. For more information on the processing of personal data, please refer to our Privacy terms

AFE944AQFP64

  • Low-power, 4-channel 24-bit ADC with integrated impedance detection ECG analog front-end

Product details

  • Parameters

  • Characteristics

  • Package

  • Description

    • 12-lead, 18-lead ECG monitor

    • 4-channel, 24-bit, SPI communication

    • -40°C to+85°C

    • • Four low-noise programmable gain amplifiers (PGA) and four high-resolution analog-to-digital converters (ADC)

    • • Low power consumption

      0.67mW per channel

    • • Input-referred noise

      6μVPP (150Hz BW,G = 6)

    • • Input bias current

      70pA

    • • Data rate

      125SPS 至 64kSPS

    • • Common Mode Rejection Ratio

      -123dB

    • • Programmable gain

      1、2、3、4、6、8、12 or 24

    • • Supports systems compliant with AAMI EC11, EC13, IEC 60601-1, IEC 60601-2-27, and IEC 60601-2-51 standards

    • • Unipolar or bipolar power supply

    • AVDD = 2.7V to 5.5V

    • DVDD = 1.65V to 3.6V

    • • Integrated right leg drive amplifier, lead-off detection, Wilson central terminal, pace detection, and test signal

    • • Integrated voltage-to-digital converter, crystal oscillator, and AC lead stimulation

    • • Digital pace detection feature

    • • Built-in oscillator and reference

    • • Serial Peripheral Interface (SPI™)-compatible serial port, daisy-chain

    • QFP64 10mm x 10mm

    • BGA64 8mm x 8mm

  • The AFE944 is a multi-channel simultaneous sampling 24-bit Δ-Σ Analog-to-Digital Converter (ADC) series, featuring a built-in Programmable Gain Amplifier (PGA), internal reference, and onboard oscillator. The AFE94x encompasses all the essential functions typically required for medical electrocardiogram (ECG) and electroencephalogram (EEG) applications. With its high level of integration and exceptional performance, the AFE94x enables the development of scalable medical instrument systems with significantly reduced size, power consumption, and overall cost. Each channel of the AFE94x includes a flexible input multiplexer (mux) that can independently connect to internally generated signals for testing, temperature monitoring, and lead-off detection. Additionally, any configured input channel can be used to generate the Right Leg Drive (RLD) output signal. The AFE94x operates at data rates up to 64 kSPS, enabling software-based pace detection. Lead-off detection is facilitated internally through pull-up/pull-down resistors or excitation current/current sources. Three integrated amplifiers are used to generate the Wilson Central Terminal (WCT) required for standard 12-lead ECG. Multiple AFE94x devices can be connected in a daisy-chain configuration in high-channel-count systems. Package options include a QFP-64 package and a compact 8mm × 8mm, 64-ball BGA. The AFE94x is rated for a temperature range of -40°C to +85°C.

Related applications

Do you need to inquire more product information?