AFE904AQFP64

  • Ultralow-power, 4-channel 24-bit ADC biopotential ECG analog front end

Product details

  • Parameters

  • Characteristics

  • Package

  • Description

    • 12-lead and 18-lead ECG medical equipment

    • 4-channel, 24-bit, SPI communication

    • -40°C to+85°C

    • • 4 low-noise programmable gain amplifiers (PGA) and 4 high-resolution analog-to-digital converters (ADC)

    • Low power consumption

      0.33mW/0.29mW per channel (low power mode)

    • • Input-referred noise

      9μVPP (150Hz BW,G = 6)

    • • Input bias current

      50pA

    • • Data rate

      62.5PS to 8kSPS

    • • Common Mode Rejection Ratio

      -105dB

    • • Programmable gain

      1、2、3、4、6、8、12 or 24

    • • Supports systems compliant with AAMI EC11, EC13, IEC 60601-1, IEC 60601-2-27, and IEC 60601-2-51 standards

    • • Unipolar or bipolar power supply

    • AVDD = 2.7V to 5.5V

    • DVDD = 1.65V to 3.6V

    • IOVDD= 1.65V to 3.6V

    • • Integrated right leg drive amplifier, lead-off detection, Wilson central terminal, pace detection, and test signal

    • • Built-in AC lead-off excitation

    • • Built-in oscillator and reference

    • • Serial Peripheral Interface (SPI™)-compatible serial port, daisy-chain

    • QFP64 10mm x 10mm

    • BGA64 8mm x 8mm

  • The AFE904 is a multi-channel, simultaneous sampling 24-bit Δ-Σ analog-to-digital converter (ADC) series, featuring a programmable gain amplifier (PGA), internal reference, and on-board oscillator. The AFE90x (1) incorporates all the functionalities typically required for medical electrocardiogram (ECG) applications. With its high integration and outstanding performance, the AFE90x enables the development of scalable medical instrument systems with significantly reduced size, power consumption, and overall cost. Each channel of the AFE90x includes a flexible input multiplexer (mux), which can be independently connected to internally generated signals for testing, temperature, and lead-off detection. Additionally, any configuration of the input channels can be used to generate the right leg drive (RLD) output signal. Lead-off detection is internally implemented through pull-up/pull-down resistors or excitation current sinks/sources. Three integrated amplifiers are utilized to generate the Wilson Central Terminal (WCT) necessary for standard 12-lead ECG. Multiple AFE90x devices can be connected in a daisy-chain configuration in high-channel-count systems. Package options include a QFP-64 package and a compact 8mm × 8mm, 64-ball BGA. The BGA version is rated for industrial temperature ranges from -40°C to +85°C.