AFE906AQFP64
-
Ultralow-power, 6-channel 24-bit ADC biopotential ECG analog front end
Product details
-
Parameters
-
Characteristics
-
Package
-
Description
-
-
12-lead and 18-lead ECG medical equipment
-
6-channel, 24-bit, SPI communication
-
-40°C to+85°C
-
-
-
• 6 low-noise programmable gain amplifiers (PGA) and 6 high-resolution analog-to-digital converters (ADC)
-
Low power consumption
0.33mW/0.29mW per channel (low power mode)
-
• Input-referred noise
9μVPP (150Hz BW,G = 6)
-
• Input bias current
50pA
-
• Data rate
62.5PS to 8kSPS
-
• Common Mode Rejection Ratio
-105dB
-
• Programmable gain
1、2、3、4、6、8、12 or 24
-
• Supports systems compliant with AAMI EC11, EC13, IEC 60601-1, IEC 60601-2-27, and IEC 60601-2-51 standards
-
• Unipolar or bipolar power supply
-
AVDD = 2.7V to 5.5V
-
DVDD = 1.65V to 3.6V
-
IOVDD= 1.65V to 3.6V
-
• Integrated right leg drive amplifier, lead-off detection, Wilson central terminal, pace detection, and test signal
-
• Built-in AC lead-off excitation
-
• Built-in oscillator and reference
-
• Serial Peripheral Interface (SPI™)-compatible serial port, daisy-chain
-
-
-
QFP64 10mm x 10mm
-
BGA64 8mm x 8mm
-
-
The AFE906 is a multi-channel, simultaneously sampling 24-bit Δ-Σ analog-to-digital converter (ADC) series, featuring an integrated programmable gain amplifier (PGA), internal reference, and on-board oscillator. The AFE90x (1) encompasses all functionalities typically required in medical electrocardiogram (ECG) applications. With its high integration and exceptional performance, the AFE90x enables the development of scalable medical instrument systems with significantly reduced size, power consumption, and overall cost. Each channel of the AFE90x is equipped with a flexible input multiplexer (mux) that can independently connect to internally generated signals for testing, temperature sensing, and lead-off detection. Furthermore, any input channel configuration can be used to generate the right leg drive (RLD) output signal. Lead-off detection can be implemented internally through pull-up/pull-down resistors or excitation current sink/source. Three integrated amplifiers are utilized to create the Wilson central terminal (WCT) necessary for standard 12-lead ECG. Multiple AFE90x devices can be cascaded in a daisy-chain configuration for high-channel-count systems. Packaging options include QFP-64 and a compact 8mm × 8mm, 64-ball BGA. The BGA version is rated for industrial temperature ranges from -40°C to +85°C.
Related applications
Related products
-
ADX920QAQFN32
Automotive-grade, low-noise PGA and 24-bit high-precision ADC analog ECG front end
-
ADX921QAQFN32
Automotive-grade, two low-noise PGAs and two 24-bit high-precision ADC analog ECG front end
-
ADX922QAQFN32
Automotive-grade integrated respiration impedance, two low-noise PGAs, and two 24-bit high-precision ADCs for an analog ECG front-end.